|
Northern Illinois University DeKalb IL
815-753-0404
Site Maintained by Michele Crase Email mcrase@niu.edu Updated 1-2008 |
Cryogenic Liquids and GasesCryogenic liquids and their boil-off vapors can rapidly freeze human tissue and may cause materials to crack or fracture under stress. All cryogenic liquids produce large volumes of gas when they vaporize (at ratios of 600:1 to 1440:1, gas: liquid) and may create oxygen-deficient conditions. Examples of common cryogenic liquids used in the labs are nitrogen oxygen, and helium. Only properly trained individuals should handle cryogenic liquids. Recommended handling practices for cryogenics:
Liquid Nitrogen Dewar Gas Release If the container is leaking Nitrogen gas via the pressure release valve do not be alarmed, this is normal operation. All of the liquid will eventually boil off and the leaking will stop. If it is leaking from any other connection close the gas valve using cryogenic gloves and notify an emergency contact listed below. The cylinders are designed to provide high pressure Nitrogen gas. The pressures will range from 250 to 350 psig. What has most likely happened is that: 1) The vacuum jacket between the inner and outer walls of the cylinder has been reduced. The vacuum acts as an insulator, higher vacuum levels result in better insulation and less gas loss. 2) The pressure release valve has frozen open. 3) The pressure building valve may be open. The proper corrective actions are to: 1) Not open the liquid valve and discharge liquid, cryogenic burns and asphyxiation can result. 2) Verify that the pressure building valve is closed. 3) If the dewar is connected to the piping system verify that the gas valve and regulator valves are open. This will relieve the excess pressure if there is an adequate gas demand from the lab. 4) Close the door to the gas room to vent the excess gas from the building. Two exhaust fans vent this room continuously.
|